Chapter 3 pH Measurement
36 HORIBA
3.1.3 Calibration Interval
Enter Calibration Interval
HINT!
We recommend performing calibration once a day, before measurement.
1. Tap on the right of the Calibration Interval
item to set the calibration interval.
2. Tap .
1. Display the Cal. Interval item when select ON at
the calibration interval. Tap on the right of
the Cal. Interval item.
2. Enter the interval day in the numerical-key
screen.
3. Tap .
The setting applies.
To cancel the settings, tap .
You can set the number of days between
calibrations to the initiate an alert.