SERVICE AND REPAIR PROCEDURES
HGB PREAMP ADJUSTMENT
4
4.29 HGB PREAMP ADJUSTMENT
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B DVM with a 3 1/2 digit display
B Pot adjustment tool or small-blade screwdriver
Procedure
1. Ensure the system is primed and that there is clear diluent in the WBC bath.
2. Open the lower chassis, as directed under Heading 4.4, and locate the Hgb Preamp card
(Figure 4.4-1).
3. Adjust the Hgb-blank reading. Using TP2 or the chassis ground for the negative lead and
TP1 for the positive lead, adjust R7 for a DVM reading of 4.5 V.
ATTENTION: When covering the photodetector, make sure no light can get to the detector.
Remove it from the WBC bath if necessary.
4. Adjust the Hgb Zero Offset. Cover the photodetector so that it receives no light and
adjust R8 until a reading of 0.0 ±1 mV is attained.
5. Restore the Hgb system and check the original Hgb-blank reading. If it has changed by
more than 0.05 V, repeat the blank and zero adjustments.
6. Restore the instrument to operating status and perform an SVP (Heading 5.1).
Note: Though a DVM is preferable, this adjustment can be performed using the service
voltage display.