PN 4237242C
5.2-2
MAINTENANCE PROCEDURES
PMI RECOMMENDATIONS
Adjustment Procedures
1. Perform the Latex Gain Adjustment procedure (Heading 4.27).
2. Perform the AIM Adjustment procedure (Heading 4.28).
3. Perform the Hgb Preamp Adjustment procedure (Heading 4.29).
Verification Procedures
1. Perform the REPRODUCIBILITY AND CARRYOVER function found in the Calibration section
of the Operator’s Guide. Both these tests must PASS.
2. Have the customer run their controls.